Step-by-step instructions for configuring the scale decimal places for a test methods on a J157, J257, J357, J457 Series 5 (serial number 16x to 17x).
- Select Menu
- Enter passcode 123
- Select Add/Edit Methods
- Highlight the method you wish to edit
- Select Edit
- Select Display Configuration
- Change the value on the Decimal Places Scale to the desired value (be sure the Extra Resolution) option is checked.
- Select Save
- Press the white Home icon located on the top right corner of the window
- Confirm the change on the main screen by running a sample.