Density Meter Backup & Restore
Steps to perform a density meter backup, restore, and air water calibration. This resolves most issues when the measured value is low.
Backup & Restore
- Select Menu, if prompted for a passcode, enter 123
- Select Calibration
- Select Backup/Restore Calibration
- Select Restore Calibration Records
- From the list or calibration records, select the one that says Factory with the most recent date
- Select Restore
- Allow the restore to complete
Air & Water Calibration
- Clean and dry the tube
- Select Menu if prompted for a passcode, enter 123
- Select Calibration
- Select Calibration Adjustment
- Select Air, Water
- Select Perform
- Select Yes, then OK
- Follow the steps displayed on the screen for the Air calibration
- A window will display the results, select Accept
- Follow the steps displayed on the screen for the Water calibration
- A window will display the results, select Accept
- The final calibration window will appear, take a picture of this screen for review if you still have incorrect readings during the next step
- Perform a water measure test and record the results
- Clean and dry the tube and run your samples, confirm if the measurement is withing specifications