Density Meter Reboots While connected to AutoFlex
What to Look for When This Issue Occurs with Automation
If the power supply is starting to degrade, during density or refractometer temperature changes, the density meter could start to reset continuously. If this happens, you should either replace the power supply, the part number is below or have the density meter power come from a dedicated power supply separate from the AutoFlex system.
If replacement parts are needed, please use the following:
- P7816 – Power Cord (USA Version) – Qty 1
- P29647 – External Power Supply, 24V – Qty 1
Important: This power supply is only compatible with S4 DDM instruments.