Density Meter Reboots While in Use
What to Look for When This Issue Occurs with Automation
There is a known issue with the new AutoFlex power supply. We have confirmed that when the instrument is ramping temperature, the power supply may begin flickering, causing the DDM to enter a reboot cycle until the temperature stabilizes.
Please advise the customer or distributor to connect the density meter to a dedicated power supply separate from the AutoFlex system.
If replacement parts are needed, please use the following:
- P7816 – Power Cord (USA Version) – Qty 1
- P29647 – External Power Supply, 24V – Qty 1
Important: This power supply is only compatible with S4 DDM instruments.
If you require additional information or assistance, please contact your distributor or reach out to Rudolph Research Analytical Technical Support. We will be happy to provide further guidance and support as needed.