Step-by-step instructions for calibrating the touch screen on an APII
Follow these steps to calibrate the touch screen on an APII.
- Select the Menu option
- Enter the passcode 123 then press the Login option
- Select the Calibration and Service option
- Enter the passcode 519 then press the Login option
- Select the Instrument Settings option
- Select the Control Panel option
- Select the Turn Disk Protection Off option.
- The instrument will reboot
- Perform steps 1 - 6 again
- Select the Touch Panel Calibration option
- Select the Tools tab (if using a mouse double click. If not using a mouse touch the tab twice).
- The screen will display a red circle and a counter. Touch and hold the red circle until the counter is complete and the red circle moves to the next calibration point of the screen.
- Select OK
- Select OK
- Close the Control Panel window by pressing the red X on the upper right side of the window.
- Select the Turn Disk Protection On option
- Select Yes
- Select OK
- The instrument will reboot.
- Test the touch glass to confirm if the calibration resolved the issue.