How to perform an air, water calibration on a density meter.
Step-by-step density meter air, water calibration
Air & Water Calibration
- Clean and dry the tube
- Select Menu if prompted for a passcode, enter 123
- Select Calibration
- Select Calibration Adjustment
- Select Air, Water
- Select Perform
- Select Yes, then OK
- Follow the steps displayed on the screen for the Air calibration
- A window will display the results, select Accept
- Follow the steps displayed on the screen for the Water calibration
- A window will display the results, select Accept
- The final calibration window will appear, take a picture of this screen for review if you still have incorrect readings during the next step
- Perform a water measure test and record the results
- Clean and dry the tube and run your samples, confirm if the measurement is withing specifications