QCP Lifter Arm Calibration
Steps to configure the lifter arms on a APV Plus VI S4
QCP Lifter Calibration: *only for APV-Plus, VI S4
- Go to terminal type “qpcal”
- Type “h” home to set calibration to 0
- Shift + “z” to increase calibration until both lifter arm is in downward position and set to 22,000
- Press “c” to calibrate
- Press “3” left arm goes up
- Press “0” should be aligned. Should be sitting firmly on the rail
- Press “1” left arm goes down
- Press “2” both arms go down
- Press “c” to calibrate
- When complete and calibrated type “a” to accept (save)