Performing a factory restore on a J57 5 series refractometer.
Step-by-step instructions for restoring the instruments calibration to factory default.
Before starting, make sure the prism is clean and dry.
- Select Menu
- Enter passcode 123
- Select Enter
- Select Setup
- Select Service Calibration
- Enter the passcode 519
- Select Reload Factory Settings
- Select Yes
- Select Yes
- Select Continue
- Select Home
- Select Zero, follow the test and provide results.