Step-by-step instructions to perform a preventive maintenance on a J275/357/457 Refractometer.
Guidelines for Executing Preventive Maintenance in Accordance with the User Manual
- Remove any contaminants from the equipment by using Windex or alcohol to clean off any foreign material.
- Inspect and replace all consumables.
-Fan Filter P20018.
-Desiccant P20245.
- Check LED References (Section 6 from the User Manual A24145 Rev O).
- Reset the M and B (Refer to Pg. 90 from the User Manual A24145 Rev O).
- Reset Factory Settings or QMC. (Refer to Page 91 from the User Manual A24145 Rev O).
- Perform 2-point calibration. (Refer to page 80 from the User Manual A24145 Rev O).
- Ensure that the calibration is saved in the Factory Settings. (Refer to page 198 from the User Manual A214145 Rev O).
- Complete the Certified Fluids Kit testing and document the outcomes for each calibration fluid.