Video camera is not visible

There may be an issue where the u-tube camera/video option is not available. Follow these steps to launch the dpinst file.

Step 1 – Getting Ready

Connect a keyboard and mouse to the density meter.

Step 2 – Turn Disk Protection Off

  1. a) Select Menu – if prompted, enter the passcode 123
  2. b) Select Operational Parameters
  3. c) Select Instrument Configuration
  4. d) Select Control Panel
  5. e) Select Turn Disk Protection Off
  6. f) Select Yes (the instrument will reboot)

Step 3

  1. a) Select Menu – if prompted, enter the passcode 123
  2. b) Select Operational Parameters
  3. c) Select Instrument Configuration
  4. d) Select Control Panel
  5. e) Select Windows Explorer
  6. f) Select Local Disk (C:)
  7. g) Select and open the Camera Settings folder
  8. h) Open the Videology folder
  9. i) double click on the dpinst file
  10. j) follow the steps on the window
  11. k) close the open windows by clicking on the red X located on the upper right corner
  12. l) go back to the main user interface window
  13. m) restart the instrument

Step 4 – Turn Disk Protection On

  1. a) Select Menu – if prompted, enter the passcode 123
  2. b) Select Operational Parameters
  3. c) Select Instrument Configuration
  4. d) Select Control Panel
  5. e) Select Turn Disk Protection Off
  6. f) Select Yes (the instrument will reboot)