Video camera is not visible
There may be an issue where the u-tube camera/video option is not available. Follow these steps to launch the dpinst file.
Step 1 – Getting Ready
Connect a keyboard and mouse to the density meter.
Step 2 – Turn Disk Protection Off
- a) Select Menu – if prompted, enter the passcode 123
- b) Select Operational Parameters
- c) Select Instrument Configuration
- d) Select Control Panel
- e) Select Turn Disk Protection Off
- f) Select Yes (the instrument will reboot)
Step 3
- a) Select Menu – if prompted, enter the passcode 123
- b) Select Operational Parameters
- c) Select Instrument Configuration
- d) Select Control Panel
- e) Select Windows Explorer
- f) Select Local Disk (C:)
- g) Select and open the Camera Settings folder
- h) Open the Videology folder
- i) double click on the dpinst file
- j) follow the steps on the window
- k) close the open windows by clicking on the red X located on the upper right corner
- l) go back to the main user interface window
- m) restart the instrument
Step 4 – Turn Disk Protection On
- a) Select Menu – if prompted, enter the passcode 123
- b) Select Operational Parameters
- c) Select Instrument Configuration
- d) Select Control Panel
- e) Select Turn Disk Protection Off
- f) Select Yes (the instrument will reboot)